| Spectrum
ANALYSIS |
Real ANALYSYS:
Professional RF measurement devices use a frequency
dependant measurement approach, the so-called
spectrum analysis. In a certain frequency range, the
individuals signals and their respective strengths are
being broken down, for example into a "bargraph" display
(see SPECTRAN® screenshots on the right). The height of
the individual bars represents the corresponding signal
strength. For the 3 strongest signal sources, SPECTRAN®
automatically displays the exact frequency and signal
level, thanks to its "Auto Marker" feature. Of course, you
can also setup the filter width and the frequency range to
be analysed as you like.
In the EMF (LF) spectrum shown here, a frequency range
of approx. 20Hz to 60Hz from left to right is being
analysed. During analysis, the Auto Marker feature has
determined - fully automatic - two main signal sources:
Signal#1=30Hz at 45µT
Signal#2=50Hz (mains power) at
75µT
In the RF spectrum shown, a frequency range of approx.
100MHz to 7GHz from left to right is being analysed (full
sweep). During analysis, the Auto Marker feature has
determined - fully automatic - three main signal
sources:
Signal#1=942MHz (GSM communications) at -63dBm
Signal#2=2024MHz (UMTS) at -23dBm
Signal#3=5832MHz (802.11a WLan) at
-42dBm
Thanks to its DIRECT frequency display of the
individual signal sources, a mapping of measurement results
to the corresponding radiation sources is
possible.
|

Spectrum display and automatic multi-marker display on
the digital screen of a NF-SPECTRAN®
(Screenshot)

Spectrum display and automatic triple multi-marker
display on the digital screen of a HF-SPECTRAN®
(Screenshot)

Well visible: "Frequency hopping" of a DECT portable
phone between 1890 and 1900 MHz
(Screenshot)
|
NOTE:
The very common low cost BROADBAND MEASUREMENT DEVICES
(especially common among novice users) do not give you too
valuable information at all. You do receive a measurement
result (for example, 500nT), but what is the frequency ? Of
how many radiation sources is this result composed? What
causes the radiation ? Are there multiple sources ? Is the
reading maybe even composed of several signal sources ? What
is the corresponding exposure limit for this reading ? What
IS the exposure limit if a mixture of several frequencies is
present ? Questions and yet more questions WITHOUT an answer.
In contrast, using a SPECTRUM ANALYSER, you will ALSO
consider the frequency of the signal sources, as the
frequency is JUST as important as the signal strength itself.
ONLY using the frequency information, a radiation source can
be isolated reliably, and ONLY using the frequency,
calculation of frequency-dependant exposure limits can be
accomplished.
Due to the MISSING frequency reference with low cost
broadband units, DRAMATIC misinterpretations can arise: when
exposure limits are calculated, these can be wrong by MORE
than a factor of 1000!! (this is NOT an error, we do really
mean ONE THOUSAND).
|
| LONG-TERM MEASUREMENT (Data
logging feature) |
For SERIOUS
measurements:
SPECTRAN® measurement devices with data logger allow
long-term recordings of measurement results over a
freely adjustable period of time. This is particularly
indispensable for SERIOUS evaluation of exposure by
appliances and machinery which have a changing power
consumption or radiation strength over time. Examples for
these include railroads, power lines and plants, but also
home appliances and their respective power cables, and
various high-frequency transmission facilities like mobile
phone transmission towers, mobile phones, radar etc.
Depending on the time of day, CONSIDERABLE variation of
exposure can occur (see attached graphics). WITHOUT long-term
recordings, MASSIVE misinterpretation of total exposure can
occur. With long-term data logging using SPECTRAN®, the
daily variation of exposure can be recorded and analysed.
Thus, the ACTUAL total exposure can be evaluated
precisely.
With this functionality, you can even discover sporadic EMC
problems which would otherwise be very hard to detect.
Even though SPECTRAN® units "only" last 4 or 7 (depending
on model) hours with one battery charge, the intelligent
"Powerdown mode" enables much longer data logging and
measurement timespans. Finally, if this is not enough, the
external power supply can be used to extend the recording
timespan infinitely. |

Daily variation of various radiation sources discloses
MASSIVE variation in exposure

Daily variation of this RF transmitter discloses EXTREME
variation in time
|
NOTE:
Long-term recording with SPECTRAN® units is particularly
INDISPENSIBLE with measurements of appliances and their
cabling if their power output or radiation varies
considerably with time. The attached screenshots show the
EXTREME variations which can occur in the course of a day.
This is a fact widely unknown by many users. In the worst
case, measured at the "wrong" time of day, very low levels
could be measured, even though the exposure, not to talk
about peak exposure, might be higher by a factor of ONE
HUNDRED THOUSAND (this is NOT an error) at a different time
of day. Great caution needs to be taken, especially if
measurement is performed as a commercial service: The very
common practice of short-term measurements (being more
cost-effective) often cannot evaluate the ACTUAL daily level
of exposure.
|
| EXPOSURE LIMITS |
At the push of
a button:
Exposure limit calculation used to be a complex and awkward
procedure even for the professional, as most of the time, a
chaotic mixture of an abundance of different frequencies,
modulations and signal strengths is present.
The indispensible, highly complex calculation of
frequency-dependant exposure limits can ONLY be performed
CONFORMING TO STANDARDS by a spectrum analyser with
high-performance software. Not a problem for SPECTRAN®
units: They can calculate even several authoritative exposure
limits, precautionary limits and recommendations (simply
selectable via a button) and display these as a practical
bargraph display (including convergence display in
percent!), while the measurement is running.
The attached SPECTRAN® screenshot demonstrates how it
works: At the push of a button, the ICNIRP exposure limit has
been chosen among the various available exposure limits.
SPECTRAN® now automatically calculates convergence of
this limit. For achieving this, often thousands of complex
calculations have to be performed per second, and a steady
scan of the entire frequency range needs to be performed. A
true nightmare for every processor. In our test case, the
graphic display shows an approximation towards the ICNIRP
limit by 6,06%. If you use a HF-6080 or NF-5030 you can even
cover the total ICNIRP-banwidth (depending on frequency).
Hence, even the novice can perform exposure limit
calculations ACCORDING TO STANDARDS without having to use
complex tables and calculators.
It really can't get any simpler. |

Graphic display of frequency-dependant exposure limits.
They disclose the INDISPENSIBLE consideration of signal
frequency
SPECTRAN® displays exposure limits both as
percentage as well as a bargraph display. Our example shows
approximation to the ICNIRP exposure limits by 6,06%.
(Screenshot)
|
NOTE:
The very common low cost BROADBAND MEASUREMENT DEVICES
(especially common among novice users) do not give you too
valuable information at all. Due to the MISSING frequency
reference with low cost broadband units, DRAMATIC
misinterpretations can arise: when exposure limits are
calculated, these can be wrong by MORE than a factor of
1000!! (this is NOT an error, we do really mean ONE
THOUSAND).
|
| The new standard: 3D
MEASUREMENT |
3D magnetic
field measurement:
Mismeasurement caused by wrongly adjusting the measurement
device in space or troublesome and complex 3D calculations
with a calculator are a problem of the past from now on,
thanks to SPECTRAN® EMF (LF) measurement devices. All
SPECTRAN® EMF measurement devices can measure magnetic
fields directly in 3D! Starting with the SPECTRAN®
NF-1010E, field strengths of the individual X, Y and Z axes
can even be shown seperately. This has become possible thanks
to the newest development from the Aaronia laboratories: Our
high-tech REAL 3D miniature sensor coil. Consisting of a
specially crafted nylon base with 3 independant windings made
of ultra-thin, 0,05 mm! wire, it impresses with its extremely
high sensitivity. It allows measurement of magnetic fields in
all 3 spacial dimensions. The signal processor (DSP) of the
SPECTRAN® performs the resulting highly complex
calculations. You receive perfect 3D measurement results
which can otherwise only be achieved by using highly
professional equipment. |

Aaronia REAL-3D magnetic field sensor
|
NOTE:
Units commonly found on the market which are particularly in
use by novices for "EMF measurement" can, for the most part,
ONLY measure magnetic fields in ONE dimension. Hence, by
misaligning the measurement device in space, SERIOUS
mismeasurement can occur: in extreme cases, no magnetic
field might be shown at all even though an extremely high
exposure is present. Tedious and complex 3D measurements are
often not performed at all, or wrongly (especially by the
novice). In contrast, SPECTRAN® can performs magnetic
field measurements in 3D and thus offers PROFESSIONAL
measurement right from the start!
|
| PROFESSIONAL PC analysis
software included for free |
Simply AMAZING:
The PROFESSIONAL PC analysis software demonstrates
SPECTRAN's vast capabilities. This software can be used IN
ADDITION to SPECTRAN and offers an incredible amount of
features. All this for FREE. Just download it from our
homepage, and your PC turns into a real spectrum analyser
with a huge display:
- MULTI-device capability!!! Remote control of
SEVERAL SPECTRAN units. These can be controlled and their
data displayed AT ONCE on a single
PC.
- HIGH-RESOLUTION!, freely scalable, coloured
spectrum display with falloff
function..
- Display of CHANNEL IDENTIFIERS!!! for EXACT
identification of providers. Channel numbers etc. freely
programmable and extensible!
- Up to 10! markers with frequency and level
display.
- Intuitive zoom control with very comfortable
frequency adjustment.
- High quality "waterfall"-display with TIMECODE.
Colour scale freely configurable. Size freely scalable.
Optional display of data DIRECTLY ON TOP OF THE GRAPH by
pointing with your mouse and
CTRL-clicking!
- High-resoution SLOT ANALYSER with 3D
display!!
- SUPER-LOGGER: ALL data can be written to disk
continuously. File format is readable by spreadsheet
applications, for creating custom reports,
etc.
- Freely positionable windows for comfortable entry
of frequency, RBW, sweeptime etc.
etc.
- Various pre-defined profiles for DECT, UMTS, GSM,
WLan etc. etc. for instant recall. Incl. optimal
parameters and extensive channel information! Freely
programmable and extensible!
- Independant main display with SIMULTANEOUS display
of dBm, dBµV, V/m, W/m2 and A/m, each
with AUTORANGE. Freely transposable and scalable.
- SUPERB exposure limit display with various
profiles (ICNIRP, Salzburg precautionary values, ECOLOG,
etc. etc.). Freely programmable with a virtually infinite
amount of display options.
- Functionality to update SPECTRAN measurement
device firmwares.
- Freely programmable key assignments and labels for
SPECTRAN measurement devices.
- Filemanager and COMPILER for creation and
management of YOUR OWN PROGRAMS for SPECTRAN measurement
devices.
- "Rename" option for renaming any of your SPECTRAN
units (for example, including location) for better
identification
- etc. etc. etc.
|


AMAZING: The PROFESSIONAL PC software for SPECTRAN. Get to
know SPECTRAN's real capabilities!
|
| HUGE digital LC
display |
Pure luxury:
The huge, high resoution digital LC display more than 80x60
mm in size! It has been developed especially for
SPECTRAN® by Aaronia. Needless to say, is uses FSTN
quality for crystal clear display. It covers a vast amount
of display capabilities (listed from top to
bottom):
- Big graphics display (pixel display) with 51x25
pixels (can be used for complex spectrum and exposure
limit display, convenient menu navigation,
etc.)
- HUGE numeric display (4 digits, max.
9999)
- Large, high-resolution bargraph (50 segments) for
very responsive trend display
- Various function and mode indication display
blocks (AUDIO, MODE, MEMORY etc.)
- Large multi-function text display (6 text fields /
e.g. for displaying units, service information,
etc.)
- Three further big numeric displays (4 digits each,
each max. 9999 / e.g. for simultaneous display of MIN,
MAX, AVG values, 3 marker values,
etc.)
- Triple multi-function text displays (Next to each
of the previously described numeric displays / 3x3 text
fields each / used e.g. for unit display for markers,
etc.)
- Three more multi-function text displays (directly
beneith the previously described displays / 3x6 text
fields each / e.g. ADDITIONAL values AND unit display for
markers, etc.)
- Battery status and charging display (incl.
"CHARGING" and "LOW")
- 6 numerical displays for flexible bargraph scaling
(incl. multiplication factor and percentage
display)
Thanks to the vast amount of flexible display options,
this device can display various readings, a bargraph AND a
spectral display (starting from NF-1010E resp. HF-2025E)
SIMULTANEOUSLY, without bothersome switching back and
forth.
Using SPECTRAN®, the user is supplied with
virtually ALL information AT ONCE, without the hassle of
switching back and forth. Where else can you find such
luxury at a similar price ?
|

The COMPLETE digital multifunction display of
SPECTRAN® at a glance
|
Overview of features SPECTRAN®
RF Spectrum Analyzer APPLICATION EXAMPLES: Measurement of
(active) radar, mobile communications, mobile phones, UMTS,
DECT phones, transmission towers, WLan, Wifi, Bluetooth,
microwave ovens, amateur radio, TETRA, radio/TV broadcast
stations, etc.
|
| |
HF-2025 |
HF-2025E |
HF-4040 |
HF-4060 |
HF-6060
V4 |
HF-6080
V4 |
HF-60100
V4 |
| SPECIFICATIONS base unit* |
For the NOVICE |
For the INTERMEDIATE |
For the PROFESSIONAL |
| Frequency range Min |
700MHz |
700MHz |
100MHz |
100MHz |
10MHz |
10MHz |
1MHz |
| Frequency range Max |
2,5GHz |
2,5GHz |
4GHz |
6GHz |
6GHz |
8GHz |
9,4GHz |
| Optional PEAK Power-Detector (Maximum usable
frequency)*** |
- |
2,5GHz |
4GHz |
6GHz |
6GHz |
8GHz |
10GHz |
| AVG
Noise Level (1Hz) |
-80dBm |
-80dBm |
-90dBm |
-90dBm |
-135dBm |
-145dBm |
-155dBm |
| AVG
Noise Level (1Hz) with PreAmp |
- |
- |
- |
- |
-150dBm |
-160dBm |
-170dBm |
| Maximum Level |
0dBm |
0dBm |
0dBm |
0dBm |
+10dBm |
+10dBm |
+40dBm** |
| Filter bandwith (RBW) Min |
1MHz |
1MHz |
100kHz |
100kHz |
10kHz |
3kHz |
1kHz |
| Filter bandwith (RBW) Max |
50MHz |
50MHz |
50MHz |
50MHz |
50MHz |
50MHz |
50MHz |
| EMC-Filter (RBW) 9kHz, 120kHz, 5MHz; 20MHz;
40MHz |
- |
- |
- |
- |
Yes |
Yes |
Yes |
| Accuracy Base unit (typical) |
+/-4dB |
+/-4dB |
+/-3dB |
+/-3dB |
+/-2dB |
+/-2dB |
+/-1dB |
| Vector power measurement (I/Q) and True
RMS |
- |
- |
Yes |
Yes |
Yes |
Yes |
Yes |
| Lowest possible SampleTime |
100mS |
100mS |
100mS |
100mS |
1mS |
1mS |
1mS |
| FEATURES |
| 14Bit Dual-ADC &
DDC-Hardware-Filter |
- |
- |
- |
- |
Yes |
Yes |
Yes |
| Standards-conformant exposure limits
(ICNIRP, BGV B11, BlmSchV etc.) |
- |
Yes |
Yes |
Yes |
Yes |
Yes |
Yes |
| Extended full ICNIRP range |
- |
- |
- |
- |
- |
- |
Yes |
| Fast
ZERO-SPAN sweep |
- |
- |
Yes |
Yes |
Yes |
Yes |
Yes |
| PULS
mode |
Yes |
Yes |
Yes |
Yes |
Yes |
Yes |
Yes |
| ADVANCED HOLD mode (HOLD function) |
- |
- |
Yes |
Yes |
Yes |
Yes |
Yes |
| INTERNAL Data Logger (long-term
measurements) |
- |
- |
Yes |
Yes |
Yes |
Yes |
Yes |
| TIME-SLOT-Analyzer |
- |
Yes |
Yes |
Yes |
Yes |
Yes |
Yes |
| Internal speaker |
Piezo |
Piezo |
Yes |
Yes |
Yes |
Yes |
Yes |
| Configurable antenna and cable calibration
data |
- |
- |
Yes |
Yes |
Yes |
Yes |
Yes |
| Audio demodulation |
AM |
AM |
AM&FM |
AM&FM |
AM&FM |
AM&FM |
AM&FM |
| DISPLAY |
| DIRECT RF spectrum display |
- |
Yes |
Yes |
Yes |
Yes |
Yes |
Yes |
| Exposure limits display with simultaneous
percentage display |
Yes |
Yes |
Yes |
Yes |
Yes |
Yes |
Yes |
| Main
display in dBm, V/m, A/m or dBìV
(switchable) |
- |
Yes |
Yes |
Yes |
Yes |
Yes |
Yes |
| ADDITIONAL display in W/m² with AUTORANGE
(pW, µW etc.) |
- |
Yes |
Yes |
Yes |
Yes |
Yes |
Yes |
| High-resolution bargraph (trend
display) |
- |
Yes |
Yes |
Yes |
Yes |
Yes |
Yes |
| 3fold
marker display (ex. 3x power & frequency at
once) |
- |
Yes |
Yes |
Yes |
Yes |
Yes |
Yes |
| INTERFACES / CONNECTORS |
| Fast
USB 2.0 Interface (PC connection) |
Yes |
Yes |
Yes |
Yes |
Yes |
Yes |
Yes |
| Audio
output (2,5mm MONO) |
Yes |
Yes |
Yes |
Yes |
Yes |
Yes |
Yes |
| DC
input (max. 15V) for external power supply |
Yes |
Yes |
Yes |
Yes |
Yes |
Yes |
Yes |
| 50
Ohm SMA RF input (F) |
Yes |
Yes |
Yes |
Yes |
Yes |
Yes |
Yes |
| Jog
Dial (multi-function dial) for “one-hand
operation” |
- |
Yes |
Yes |
Yes |
Yes |
Yes |
Yes |
| OPTIONS (extra charge) |
| Option 001 (1MB memory
expansion) |
- |
- |
- |
Yes |
Yes |
Yes |
Yes |
| Option 002 (high sensitive 0,5ppm TCXO
timebase) |
- |
- |
- |
- |
- |
- |
Yes |
| Option 020 (internal, switchable 15dB
PreAmplifier) |
- |
- |
- |
- |
Yes |
Yes |
Yes |
| Option 20x (REALTIME broad band
Power-Meter) |
- |
2,5GHz |
4GHz |
6GHz |
6GHz |
8GHz |
10GHz |
| INCLUDED ACCESSORIES in addition to
the base unit |
| Miniature SMA rod antenna |
Yes |
Yes |
Yes |
Yes |
Yes |
Yes |
Yes |
| HyperLOG EMC directional LogPer antenna
(model) |
7025 |
7025 |
7040 |
7060 |
7060 |
6080 |
60100 |
| Aaronia 7,2V high-performance battery (1300mAh) +
charger |
Yes |
Yes |
Yes |
Yes |
Yes |
Yes |
Yes |
| Aluminum design transport case |
Yes |
Yes |
Yes |
Yes |
Yes |
Yes |
Yes |
| PROFI
PC analysis software (Windows,
downloadable) |
Yes |
Yes |
Yes |
Yes |
Yes |
Yes |
Yes |
| List price plus shipping |
€ 249,95 |
€ 299,95 |
€ 499,95 |
€ 599,95 |
€ 999,95 |
€ 1298,00 |
€ 1498,00 |
*Further
REALTIME spectrum analysers up to 18GHz are already in
development. Please contact us for further
details!
Preliminary specifications as of 05.12.2008. The V4 series is
available with latest Beta-Firmware. All options are
available for this series too. The Beta-Firmware is in
continuous development. Some functionality may still be
limited and not fully to specifications (Beta-Status). By
regularly checking our homepage for updates, you can always
keep your measurement device up-to-date. As soon as V1.0 of
the firmware is released, all functionality and features will
be fully available.
Range, sensetivity and accuracy can change depending on
frequency, antenna and used parameters. Precision values are
based on Aaronia calibration-reference under specific test
conditions. Unless otherwise stated, these specifications
apply for the reference condition: ambient temperature
22±3°C, relative air humidity 40% to 60%,
continuous wave signal (CW), RMS detection. V4 Noise Level
@5,555GHz. Maximum sensitivity of Rev.3 units: -90dBm
@2,2GHz.
** Internal: +20dBm. External (with optional 20dB precision
attenuator): +40dBm
*** Depending on frequency the optional PEAK power meter
offers sensetivity up to -50dBm and max. +10dBm input power
with an extremely fast response time.
© Aaronia AG, DE-54597 Euscheid, Germany,
www.spectran.com, Phone. ++49(0)6556-93033
|
Overview of features SPECTRAN® EMF (NF)
Spectrum Analyzer APPLICATION
EXAMPLES: Measurement of traction power, power lines,
power cables, lamps, power supplies, transformer stations,
various appliances in households, offices and
manufacturing
|
| SPECIFICATIONS base
unit* |
For the
NOVICE |
For the
INTERMEDIATE |
For the PROFESSIONAL |
| |
NF-1010* |
NF-1010E* |
NF-3010* |
NF-3020* |
NF-5010* |
NF-5030* |
| Frequency range
Min |
10Hz |
10Hz |
10Hz |
10Hz |
1Hz |
1Hz |
| Frequency range
Max |
2kHz |
10kHz |
100kHz |
400kHz |
1MHz |
30MHz** |
| Range electrical field
[V/m] (typical) Min (1D) |
1V/m |
1V/m |
0,1V/m |
0,1V/m |
0,1V/m |
0,1V/m |
| Range electrical field
[V/m] (typical) Max (1D) |
2kV/m |
2kV/m |
5kV/m |
5kV/m |
5kV/m |
20kV/m |
| Range magnetic field
[Tesla] (typical) Min (3D!) |
10nT |
10nT |
1nT |
1nT |
1nT |
1pT** |
| Range magnetic field
[Tesla] (typical) Max (3D!) |
100µT |
100µT |
100µT |
100µT |
100µT |
2mT |
| Range magnetic field
[Gauss] (typical) Min (3D!) |
100µG |
100µG |
10µG |
10µG |
10µG |
10nG** |
| Range magnetic field
[Gauss] (typical) Max (3D!) |
1G |
1G |
1G |
1G |
1G |
20G |
| Filter bandwidth
Min |
5Hz |
5Hz |
1Hz |
1Hz |
1Hz |
1Hz |
| Filter bandwidth
Max |
10kHz |
100kHz |
300kHz |
300kHz |
1MHz |
1MHz |
| Accuracy base unit
(typical) |
5% |
5% |
5% |
3% |
3% |
3% |
| FFT (Resolution in
points) |
64 |
64 |
64 |
64 |
1024 |
1024 |
| Vector power
measurement (I/Q) and True RMS |
- |
- |
 |
 |
 |
 |
| FEATURES |
| Standards conformant
exp. limits (ICNIRP, BGV B11, BlmSchV etc.) |
- |
 |
 |
 |
 |
 |
| Extended full ICNIRP
range |
- |
- |
- |
- |
- |
 |
| Isotropic (3D) AC
magnetic field measurement |
 |
 |
 |
 |
 |
 |
| Supports custom P-Code
software |
- |
- |
 |
 |
 |
 |
| ADVANCED HOLD mode
(HOLD function) |
- |
 |
 |
 |
 |
 |
| INTERNAL data logger
(long-term measurements) |
- |
- |
 |
 |
 |
 |
| FLASH memory including
firmware update (over the Internet) |
- |
16K |
64K |
64K |
64K |
64K |
| “Clear
text” signal identification with direct frequency
display |
- |
 |
 |
 |
 |
 |
| Integrated battery
charging circuitry |
- |
 |
 |
 |
 |
 |
| Internal
speaker |
Piezo |
Piezo |
 |
 |
 |
 |
| Audio
demodulation |
AM |
AM |
AM |
AM |
AM&FM |
AM&FM |
| DISPLAY |
| Fast FFT or DFT
spectrum analyses |
- |
 |
 |
 |
 |
 |
| Limit
calculation with simultaneous percentage
display |
 |
 |
 |
 |
 |
 |
| X, Y, Z Axis display or
Vectorproduct (only M.-Field) |
- |
 |
 |
 |
 |
 |
| Main display in V/m,
Tesla, Gauss or A/m (switchable) |
- |
 |
 |
 |
 |
 |
| High-resolution 50
segment bargraph (trend display) |
 |
 |
 |
 |
 |
 |
| 3fold marker display
(ex. 3x field strength & frequency at once) |
- |
 |
 |
 |
 |
 |
| INTERFACES /
CONNECTORS |
| Fast USB 2.0 interface
(computer connection) |
- |
 |
 |
 |
 |
 |
| Audio
output |
 |
 |
 |
 |
 |
 |
| DC input (max. 15V) for
external power supply |
 |
 |
 |
 |
 |
 |
| External ultra
sensetive signal input (SMA input) with max. 0,2V |
- |
- |
- |
- |
 |
 |
| Jog Dial
(Multi-functional dial) for "one-hand operation" |
- |
- |
 |
 |
 |
 |
| OPTIONS (extra
charge) |
| Option 001 (1MB
memory expansion) |
- |
- |
- |
- |
 |
 |
| Option 005
(12Bit DDC / offers ultra high sensetivity up
to 1pT) |
- |
- |
- |
- |
- |
 |
| Option 006
(Measure 3D static magnetic fields)* |
- |
- |
- |
- |
- |
 |
| Option 009
(Ultra high 24Bit resolution on static magnetic
fields) |
- |
- |
- |
- |
- |
 |
| Option 010
(Expanded frequency range up to 30MHz e.g.
RFID) |
- |
- |
- |
- |
- |
 |
| INCLUDED
ACCESSORIES in addition to the base unit |
| Aaronia 7,2V
high-performance battery (1300mAh) + charger |
- |
 |
 |
 |
 |
 |
| Aluminum design
transport case incl. padding inlays |
opt. |
 |
 |
 |
 |
 |
| PROFESSIONAL PC
analysis software (Windows, download via
Internet) |
- |
 |
 |
 |
 |
 |
| List
price plus shipping |
€ 299,95 |
€ 349,95 |
€ 399,95 |
€ 449,95 |
€ 549,95 |
€ 599,95 |
| Online Shop: |
|
|
|
|
|
|
*Preliminary specifications
as of 12.03.2009. NF series is available with latest
BETA-Firmware. ALL options are available for this series. The
BETA firmware is in continuous development. Some
functionality may still be limited and not fully to
specifications (BETA status). By regularly checking our
homepage for updates, you can always keep your measurement
device up-to-date. As soon as version 1.0 of the firmware is
released, all functionality and features will be fully
available.
Range and accuracy can change depending on frequency, sensor
and used parameters. Precision values are based on Aaronia
calibration-reference and only valid under specific test
conditions. Unless otherwise stated, these specifications
apply for the reference condition: ambient temperature
22±3°C, relative air humidity 40% to 60%,
continuous wave signal (CW), RMS detection
Option 006 offers a range of 100µG-6G (10nT-600µT).
You can "zero" the static field sensor (Option 006) by using
our "Zero Gauss" chamber.
**Standard: 1MHz. Only with option 008 up to 30MHz. /
Standard: 1nT. Only with option 005 up to 1pT.
© Aaronia AG, DE-54597 Euscheid, Germany,
www.spectran.com, Tel. ++49(0)6556-93033
|